PSL models upgraded to RCG2.8.5. New safe.snap files tested

Peter and Dave.

We rebuilt all h1psl0 models against RCG2.8.5 (h1ioppsl0, h1psliss, h1pslfss, h1pslpmc, h1psldbb). We restarted the models, using the latest safe.snap burt restore files created on Sept 5th.

After the 14th August 2014 power outage, I have downgraded the PSL to RCG2.8.3 as part of our attemp to open the closed shutter.

We were unable to reproduce the problem of not being able to open the Beckhoff controlled shutter in the diode room. When the PMC model was started the shutter closed. On the LASER MEDM screen, we RESET the Flow Sensor External Shutter Request (required confirmation) and then pressed the Control Room RESET (and confirmed), the shutter then opened.