Reports until 11:32, Tuesday 16 September 2014
H1 AOS
alastair.heptonstall@LIGO.ORG - posted 11:32, Tuesday 16 September 2014 (13952)
H1OAF0 shutdown - causes TCS laser shutoff

Shutdown of H1OAF0 to replace TCS DAC card.  This caused TCS chiller and laser shutdown.  The laser chiller temperature set point is  controled by the DAC, and a zero volt output puts the chiller setpoint outside the safe operating range causing the chiller to shutdown. This then causes the laser to shutdown.