Reports until 15:32, Wednesday 17 September 2014
H1 TCS
patrick.thomas@LIGO.ORG - posted 15:32, Wednesday 17 September 2014 - last comment - 19:01, Thursday 18 September 2014(13977)
TCSX rotation stage not moving
Alastair, Greg, Patrick

Alastair tried to use the TCSX rotation stage to change the power this morning. Nothing on the medm screen seemed to indicate that anything was happening.

When I request a search for home, the 'Actual Cmd Position', 'Actual Velocity' and 'Drive Time (ms)' change on the 'Rotation Stage Readback' medm, but the 'Counter Actual Angle' does not.
Comments related to this report
patrick.thomas@LIGO.ORG - 16:45, Wednesday 17 September 2014 (13985)
Vern checked and says that there is no interlock plug installed.
greg.grabeel@LIGO.ORG - 17:16, Wednesday 17 September 2014 (13986)TCS
The interlock bypass plug has appeared to fix the issue. We're not sure where what was in there before, or where it had gone, but thanks to Vern we have a good solution in place. Fingers crossed this solves the issues.
greg.grabeel@LIGO.ORG - 19:01, Thursday 18 September 2014 (14024)TCS
Here are the new interlock bypass caps that Vern rigged up for us. Seems to have solved the issues with the rotation stage as no more errors popped up while doing the projection work.
Images attached to this comment