(Corey, Greg, Jim)
This balancing encompasses two days. We balanced the system by monitoring digital voltmeters connected to each Corner's mini-rack. For the Stage1 CPS's we aimed for voltages under +/- 0.200V, and for the Stage2 (Fine) CPS's we aimed for voltages under +/-0.500V. A negative voltage connotates a small gap w.r.t. Target Face, and a positive voltage obviously means a bigger gap.
On Wed, we primarily addressed Stage2, and then stopped since we didn't have enough weights, and needed to consult MIT with our progress. The notes for Wednesday's work is attached. For today's work, we focused on finishing out Stage1 Balancing. After the entire system was balanced we also noted mass placement for the system.
Corner1
Stage1
H: -0.032 V
V: -0.062
Corner2
Stage1
H: 0.016
V: -0.115
Corner3
Stage1
H: -0.021
V: 0.055
Corner1
Stage1
H: -0.062
V: -0.014
Stage2
H: 0.066
V: -0.299
Corner2
Stage1
H: 0.115
V: -0.015
Stage2
H: 0.103
V: 0.144
Corner3
Stage1
H: 0.190
V: 0.033
Stage2
H: 0.432
V: -0.380
Below are the masses used to attain the current balanced state:
Keel:
(26)D0903136
(2) -1200 Type6
Optical Table:
(6) Trim Mass Damper Assys
CORNER 1
Stage1
Wall: D0902614, -2615, -2616Type1 , -2616 Type2, -2616Type3, -2616Type5
Side Wall: -2612
Stage2
Walls (D0701200): (1)Type1, (3)Type2, (1)Type3, (3)Type4
CORNER 2
Stage1
Wall: D0902614, -2616Type1, -2616Type1, -2616Type2
Side Wall: -2612
Stage2
Walls (D0701200): (2)Type1, (1)Type4, (1)Type5
CORNER 3
Stage1
Wall: D0902614, -2615, -2616Type2, -2616Type5
Side Wall: -2612
Stage2
Walls (D0701200): (1)Type2, (1)Type6