X2 accumulated from Sept 18 through Sept 22.
N2 leak rate is <4e-9 tl/s
H2 outgassing is 5.9e-15 tl/s/cm2.
The RGA calibration is assumed to be 319 torr/amp.
No temperature data was taken for this module. My guess is the average tube temperature has been 25C.