PSL Check: 9/29/2014
Laser Status:
-
SysStat is good
-
Output power is 33.6 W (should be around 30 W)
-
FRONTEND WATCH is active
-
HPO has multiple red, LOCKED, LRA, WATCH and WARNINGS.
PMC:
-
It has been locked 2 days, 20 hr , and 56 minutes (should be days/weeks)
-
Reflected power is 2.0 Watts and PowerSum = 25.6 Watts.
-
(Reflected Power should be <= 10% of PowerSum)
FSS:
-
It has been locked for 1 min (should be days/weeks), keeps glitching.
-
Threshold on transmitted photo-detector PD = 1.753 V (should be at least 0.9V)
ISS:
-
The diffracted power is around 9.75 % (should be 5-15%)
-
Second loop is RED.
-
Last saturation event was 1 d, 23 h and 20 minutes ago (should be days/weeks)