See the attached spectra to see the obvious problem sensor. I swapped the cables from the H & V sensors to the Pier Pod and the problem signal followed the cable. So, the problem is upstream with the cable or sensor. The HAM HEPI Horizontal L4C is certainly a pain to access:
Step1-- lock the local foot (already there on HAM1)
2 -- Disconnect the HEPI Actuator--recenter and lock up.
3 -- Remove the two Actuator mounting adapters.
4 -- Remove the L4C
5 -- Replace & Level the L4C
Reverse steps and hope you didn't move the platform enough to raise Daniel's notice (we are pretty good at not moving the system during this.)