Jenne, Alexa, Sheila
Since the asc had IPC senders for signals to OM1 and OM2, but the only revciever in the h1sushtts model were from the OMC model, we added new recievers. Anamaria told us where to find the livingston model, and confirmed that it was up to date, so we copied what they had done there, just summing the inputs from the OMC model and ASC.
Complied, installed, seems to be working, and we can now use the DC signals from our AS WFS for a centering servo.