Displaying report 1-1 of 1.
Reports until 16:22, Friday 03 October 2014
H1 TCS (TCS)
greg.grabeel@LIGO.ORG - posted 16:22, Friday 03 October 2014 (14294)
TCS ITMy Rotation stage fixed?
The rotation stage on ITMy had become unresponsive last week after some cable dressing was done. Restarting Beckhoff had proven ineffective and Fil was unable to find any lose cables. Unplugging and replugging the interlock bypass on the D1300131 box. This possibly had the effect of resetting the relay that blocks communication to the relay stage. Whatever happened the rotation stage is working again.