Displaying report 1-1 of 1.
Reports until 14:16, Tuesday 17 June 2014
H1 ISC
keita.kawabe@LIGO.ORG - posted 14:16, Tuesday 17 June 2014 - last comment - 16:43, Wednesday 08 October 2014(12392)
HAM6 checked out for grounding

I and Dan replaced the bad UL BOSEM unit (SN of bobbin part of that assy is SN250, D0601106-C) with a good one (SN282) pulled out of spare "half-assembled" tip-tilt assy (base plate SN 034).

We also replaced UR one (SN258) with a spare (SN134) that was pulled out of the same half-assembled TT assy simply because we were not sure if the open voltage of SN258 (about 20000 counts) was too small. Later it turns out that 20k counts open voltage is not abnormal.

The newly installed units were of course assembled in the past but without fit check, and it turns out that the gap around the connector was non-existent and I had to loosen some screws and shift things around to make a good gap. I also used aluminum foil shim trick.

After this, we centered BOSEMs for all Tip-tilts to half the open voltage.

  UL LL UR LR
OM1 -16k -12k -16k -13k
OM2 -11k -11.5k -15.5k -16k
OM3 -16k -14k -16k -11.5k

WFS DC, AS_C QPD, OMCR QPDs, OMC QPDs and OMC DCPDs all responded to flash light.

Beam diverters moved back and forth.

Picomotors moved. On cable #234, the first, the second and the third channels correspond to WFSA, WFSB and AS_C picomotors. On cable #235, the first channel is the downstream 2" mirror for the QPD sled, the second the upstream 1" steering.

Checked ground loops from outside the chamber, and all cables including SUS were good except beam diverters (we know that they ground inside the chamber) and OMC QPDs (shield grounted to the DCPD signal ground) and OMC PZTs (same as OMC QPDs).

Comments related to this report
keita.kawabe@LIGO.ORG - 14:28, Tuesday 17 June 2014 (12393)

OMC PZT cable shielding also is grounded to DCPD signal ground.

Same as OMC QPDs. It's not clear from D1300376 (OMC PZT cable document) but there seems to be continuity from the shield to the mighty mouse connector shell, which is mounted on the aluminum bracket on the OMC.

LLO alog explaining these: https://alog.ligo-la.caltech.edu/aLOG/index.php?callRep=10121

keita.kawabe@LIGO.ORG - 14:48, Tuesday 17 June 2014 (12396)

As of now, ISI is locked. OMC is free.

keita.kawabe@LIGO.ORG - 14:32, Tuesday 17 June 2014 (12394)

Beam diverter cable sheild needs to be cut at the rack or at the chamber.

With the cables disconnected from the remote rack Beckhoff motor driver module, there's continuity between the cable shell and the rack. Relevant cables are cable #316 and #317.

keita.kawabe@LIGO.ORG - 14:35, Tuesday 17 June 2014 (12395)

PZT capacitance:

From the outside of the chamber, used a short DB25 and a break out board to check that PZT pins are not short circuited to stupid things. Since I was able to, I also measured the capacitance.

Pin1-14: 474nF.

Pin2-15: 428nF.

These of course include the in-vac cables and short DB25 cable I used outside.

koji.arai@LIGO.ORG - 14:56, Wednesday 18 June 2014 (12412)

Sounds the correct range for the PZTs. (spec: 470nF)

keita.kawabe@LIGO.ORG - 16:43, Wednesday 08 October 2014 (14376)

Correction: Corey found that the S/N of the pulled and newly put in BOSEMs are swapped in the above entry.

Bad BOSEM that has one PD pin grounded to the coil bobbin part is SN282.

  Pulled Newly installed
UL SN282, broken SN250
UR SN134, turns out that nothing was wrong. SN258