Laser Status: SysStat: Warning “VP program online” is red Output power is 33.5 W (should be around 30 W) FRONTEND WATCH is active HPO WATCH is red PMC: It has been locked 5 days, 22 hr , and 54 minutes (should be days/weeks) Reflected power is 2.0 Watts and PowerSum = 25.7 Watts. (Reflected Power should be <= 10% of PowerSum) FSS: It has been locked for 1 d, 7 h, and 9 min (should be days/weeks) Threshold on transmitted photo-detector PD = 2.26 V (should be at least 0.9V) ISS: The diffracted power is around 9.5% (should be 8-10%) Last saturation event was 2 d, 19 h and 0 minutes ago (should be days/weeks NOTE: The ISS Diffracted Power was out of its range (8-10%) this morning. It was fixed by adjusting the ISS Reference Signal value.