Reports until 10:28, Monday 13 October 2014
H1 PSL
andres.ramirez@LIGO.ORG - posted 10:28, Monday 13 October 2014 (14424)
PSL Check: 10/13/2014
Laser Status: 
SysStat: Warning “VP program online” is red
Output power is 33.5 W (should be around 30 W)
FRONTEND WATCH is active
HPO WATCH is red

PMC:
It has been locked 5 days, 22 hr , and 54 minutes (should be days/weeks)
Reflected power is 2.0 Watts and PowerSum = 25.7 Watts.
(Reflected Power should be <= 10% of PowerSum)

FSS:
It has been locked for 1 d, 7 h, and 9 min (should be days/weeks)
Threshold on transmitted photo-detector PD = 2.26 V (should be at least 0.9V)

ISS:
The diffracted power is around 9.5% (should be 8-10%) 
Last saturation event was 2 d, 19 h and 0 minutes ago (should be days/weeks


NOTE: The ISS Diffracted Power was out of its range (8-10%) this morning. It was fixed by adjusting the ISS Reference Signal value.