PSL Status
This status is reported after the adjustment/instability of the REFSIGNAL mentioned in aLog 14634.


Laser Status: 
SysStat: Warning “VP program online” is red
Output power is 29.5 W (Should be around 30 W)
FRONTEND WATCH is active
HPO WATCH is red

PMC:
It has been locked 20 days, 2 hours, 8 minutes.
Reflected power is 2.1 Watts and PowerSum = 25.7 Watts.
(Reflected Power should be <= 10% of PowerSum)

FSS:
It has been locked for 1 hour, 45 min.
Threshold on transmitted photo-detector PD = 2.27 V (should be at least 0.9V)

ISS:
The diffracted power is around 8.8% (should be 8-10%) 
Last saturation event was 1 hour, 49 minutes ago 