This status is reported after the adjustment/instability of the REFSIGNAL mentioned in aLog 14634. Laser Status: SysStat: Warning “VP program online” is red Output power is 29.5 W (Should be around 30 W) FRONTEND WATCH is active HPO WATCH is red PMC: It has been locked 20 days, 2 hours, 8 minutes. Reflected power is 2.1 Watts and PowerSum = 25.7 Watts. (Reflected Power should be <= 10% of PowerSum) FSS: It has been locked for 1 hour, 45 min. Threshold on transmitted photo-detector PD = 2.27 V (should be at least 0.9V) ISS: The diffracted power is around 8.8% (should be 8-10%) Last saturation event was 1 hour, 49 minutes ago