J. Kissel I attach pictures of the switchable configuration boards of all optical levers in the corner station. My fingers are supposedly indicative of the rack / chassis into which they're plugged, but I add a legend here anyways with the real rack numbers, and I include the exact switchable filter status. Note, I only report one of the channels' switch status, because (thank god) in every instance, all four channels are configured in the same way. Rack Board Optical Lever Config Board |------------------ Switch Status ----------------| Centered? Raw Segment Counts Number Installed? Board Switch CH Name CH Name Function HI LO ON/OFF SUS H1-R1 1 HAM2 Table? No N/A (HAM2) No 1:-339 3:-2243 (Pg 1) HAM3 Table? 4:-399 2:-3548 2 HAM2 Table? No N/A (HAM3) Maybe 1:-1260 3:-1167 HAM3 Table? 4:-1921 4:-1594 SUS H1-R2 1 none? No N/A N/A (Pg 2) 2 PR3 Yes B0 1 +24 [dB] X OFF PR3 Yes-ish 1:-5600 3:-7400 (pgs 3-5) B1 2 +12 [dB] X ON 4:-17800 2:-17500 B2 3 +6 [dB] X ON B3 4 +3 [dB] X OFF B4 5 (1:10) X ON B5 6 (1:10) X OFF B6 7 (1:10) X OFF B7 8 Latch X OFF Total PR3 TF = (1:10), G = 18 [dB] SUS H1-R4 1 HAM4 Table? No N/A (HAM4) No ADC Noise (pgs 6-7) HAM5 Table? 2 HAM4 Table? No N/A (HAM5) No ADC Noise HAM5 Table? SUS H1-R3 1 none? No N/A N/A (pg 8) 2 SR3 Yes B0 1 +24 [dB] X OFF SR3 Yes-ish 1:-9100 2:-9200 (pgs 9-10) B1 2 +12 [dB] X OFF 4:-7040 3:-9200 B2 3 +6 [dB] X OFF B3 4 +3 [dB] X OFF B4 5 (1:10) X ON B5 6 (1:10) X OFF B6 7 (1:10) X OFF B7 8 Latch X OFF Total SR3 TF = (1:10), G = 0 [dB] SUS H1-R6 1 BS Yes B0 1 +24 [dB] X OFF BS Yes-ish 1:-6800 2:-6300 (pgs 11) (pg 12) B1 2 +12 [dB] X ON 4:-7250 3:-7350 B2 3 +6 [dB] X OFF B3 4 +3 [dB] X ON B4 5 (1:10) X ON B5 6 (1:10) X OFF B6 7 (1:10) X OFF B7 8 Latch X OFF Total BS TF = (1:10), G = +15 [dB] 2 ITMY Yes B0 1 +24 [dB] X OFF ITMY Yes-ish 1:-2470 2:-2380 (pgs 15) (pg 16-20) (NOT AS SHOWN) B1 2 +12 [dB] X OFF 4:-2250 3:-2900 B2 3 +6 [dB] X OFF B3 4 +3 [dB] X OFF B4 5 (1:10) X OFF B5 6 (1:10) X OFF B6 7 (1:10) X OFF B7 8 Latch X OFF Total ITMY TF = (flat), G = 0 [dB] SUS H1-R5 1 ITMX Yes B0 1 +24 [dB] X OFF ITMX Yes-ish 1:-1000 2:-700 (pgs 13) (pg 14) B1 2 +12 [dB] X OFF 4:-1000 3:-700 B2 3 +6 [dB] X OFF B3 4 +3 [dB] X OFF B4 5 (1:10) X OFF B5 6 (1:10) X OFF B6 7 (1:10) X OFF B7 8 Latch X OFF Total ITMX TF = (1,1:10,10), G = 0 [dB] Note that I've left ITMY with all of its switches OFF because we want to assess whether the internal gain of the QPD head (see D1100290) is set to the "high" (100 [kOhm] transimpedance) or "low" (10 [kOhm] transimpedance) -- and I got booted out of the LVEA for locking the IFO. Seems like 2000 counts is a little to low, we really want something like 10000 [cts] per quadrant. We will adjust tomorrow. I've also confirmed that all over-all transfer functions for each optic has digitally compensated for correctly.