End-station Optical Lever Configuration Switch Status
J. Kissel

I've collected all the information necessary to reproduce the corner-station optical lever status report (see LHO aLOG 14749) for the end-station levers. I attach pictures, and tabulate info as before below.

Rack          Board      Optical Lever    Config Board  |------------------ Switch Status ----------------|    Centered?    Raw Segment Counts
              Number                       Installed?    Board    Switch                                        
                                                        CH Name   CH Name    Function    HI   LO    ON/OFF       

SUS H1-R2       1          ETMX               Yes          B0       1       +24 [dB]     X            OFF          No*      *SUS is Misaligned, 
  (pg 1)                 (pgs 2-3)                         B1       2       +12 [dB]     X            OFF                   so can't get info
                                                           B2       3       +6  [dB]          X       ON                
                                                           B3       4       +3  [dB]          X       ON
                                                           B4       5       (1:10)       X            OFF
                                                           B5       6       (1:10)            X       ON
                                                           B6       7       (1:10)       X            OFF
                                                           B7       8       Latch        X            OFF
                                                              Total ETMX TF = (1:10), G = +9 [dB]
 
                2          none               Yes          N/A                                                     N/A

SUS H1-R2       1          ETMY               Yes          B0       1       +24 [dB]     X            OFF          No*      *SUS is Misaligned
  (pg 4)                 (pgs 5-6)                         B1       2       +12 [dB]     X            OFF                   so can't get info
                                                           B2       3       +6  [dB]     X            OFF               
                                                           B3       4       +3  [dB]     X            OFF
                                                           B4       5       (1:10)            X       ON
                                                           B5       6       (1:10)            X       ON
                                                           B6       7       (1:10)       X            OFF
                                                           B7       8       Latch        X            OFF
                                                              Total ETMY TF = (1,1:10,10), G = 0 [dB]
 
                2          none               Yes          N/A                                                     N/A

I can confirm that any whitening stages have been correctly compensated for digitally. In addition, recall that ETMX is compensating for the gains shown above, but it's not necessary. Will fix all the levers and compensation on some (glorious) day (hopefully soon) when the IFO isn't needed.