PSL Checklist
Adjusted REFSIGNAL from -2.02V to -2.06V to bring diffracted power down from ~12% to ~10.8%.  I am hesitant to make more aggressive adjustments since last time I did, Rick was surprised that such large adjustments were required.  Further investigation?


Laser Status: 
SysStat: Warning “VP program online” is red
Output power is 29.4 W (Should be around 30 W)
FRONTEND WATCH is active
HPO WATCH is red

PMC:
It has been locked 5 days, 21 hours, 58 minutes.
Reflected power is 2.0 Watts and PowerSum = 25.4 Watts.
(Reflected Power should be <= 10% of PowerSum)

FSS:
It has been locked for 15 hour, 22 min.
Threshold on transmitted photo-detector PD = 2.20 V (should be at least 0.9V)

ISS:
The diffracted power is around 10.8% (should be 8-10%) 
Last saturation event was 1 hour, 34 minutes ago 