Adjusted REFSIGNAL from -2.02V to -2.06V to bring diffracted power down from ~12% to ~10.8%. I am hesitant to make more aggressive adjustments since last time I did, Rick was surprised that such large adjustments were required. Further investigation? Laser Status: SysStat: Warning “VP program online” is red Output power is 29.4 W (Should be around 30 W) FRONTEND WATCH is active HPO WATCH is red PMC: It has been locked 5 days, 21 hours, 58 minutes. Reflected power is 2.0 Watts and PowerSum = 25.4 Watts. (Reflected Power should be <= 10% of PowerSum) FSS: It has been locked for 15 hour, 22 min. Threshold on transmitted photo-detector PD = 2.20 V (should be at least 0.9V) ISS: The diffracted power is around 10.8% (should be 8-10%) Last saturation event was 1 hour, 34 minutes ago