Laser Status:
SysStat is good
Output power is 33.3 W
Front End WD is GREEN
HPO WATCH is RED
PMC:
It has been locked 6 day, 22 hr 10 minutes
Reflected power is 2.1 Watts and PowerSum = 23.8 Watts.
(Reflected Power should be <= 10% of PowerSum)
FSS:
It has been locked for 1 h and 57 min (should be days/weeks)?
Threshold on transmitted photo-detector PD =2.16V (should be 0.9V)
ISS:
The diffracted power was around 5.6%. The REFSIGNAL was adjusted to -02.04 to yield a diff power of ~8.2%(should be 5-15%)
Last saturation event was 2h and 3 minutes ago (should be days/weeks)