Weekly PSL Check
Laser Status: 
SysStat is good
Output power is 33.3 W (should be around 30 W)
FRONTEND WATCHdog is Active
HPO WATCH is RED

PMC:
It has been locked 13 day, 1 hr 26 minutes (should be days/weeks)
Reflected power is 2.0 Watts  and PowerSum = 25.8 Watts.
(Reflected Power should be <= 10% of PowerSum)

FSS:
It has been locked for 3 days, 21 h and 33 min (should be days/weeks)
Threshold on transmitted photo-detector PD = 2.265V (should be 0.9V)

ISS:
The diffracted power is around 7.4% (should be 5-15%)
Last saturation event was 3 days, 21 h and 38 minutes ago (should be days/weeks)