Reports until 12:10, Monday 17 November 2014
H1 SUS
sheila.dwyer@LIGO.ORG - posted 12:10, Monday 17 November 2014 (15104)
PR3 calibrations

The calibrations on PR3 angles are wrong.  In the attachecd screen shot you can see that someone moved the alignment slider -72 urad in yaw, the OSEMs sees a shift of -62 urad, and the optical lever records a shift of -26 urad

Images attached to this report