alexan.staley@LIGO.ORG - posted 19:35, Tuesday 02 December 2014 (15398)
HEPI Tidal feedback restored

Sheila, Alexa

We noticed that the HEPI tidal feedback to the ETMs was not working. We found that the ISC MON outputs were off, the ISCINF_LONG offsets were off. In addition the ETMY ISCINF gain was zero and the ETMX ISCINF gain was 1000; the nominal gains should be 1. This is now in the guardian.

Displaying report 1-1 of 1.