Displaying report 1-1 of 1.
Reports until 23:45, Tuesday 11 October 2011
H2 SUS
jeffrey.garcia@LIGO.ORG - posted 23:45, Tuesday 11 October 2011 (1548)
FMY .ini file changes

A trial transfer function measurement on the H2 FMY M1 today resulted in a stalled Matlab script when attempting to retrieve data. The issue was tracked down to a modified "H2SUSFMY.ini" file that had all but two channels in the "acquire=0" state. This is possibly due to the new FMY model installed last week with the RCG auto-generating a new .ini file.  The .ini file was edited to acquire the needed channels, and the trial measurement ran smoothly thereafter.