Laser Status: SysStat is good Output power is 33.1 W (should be around 30 W) FRONTEND WATCHdog is Active HPO WATCH is RED PMC: It has been locked 7 day, 21 hr 39 minutes (should be days/weeks) Reflected power is 2.4 Watts, and PowerSum = 25.8 Watts. (Reflected Power should be <= 10% of PowerSum) FSS: It has been locked for 0 days, 0 h and 17 min (should be days/weeks) Threshold on transmitted photo-detector PD = 1.555V (should be 0.9V) ISS: The diffracted power is around 5.700% (should be 5-15%) Last saturation event was 0 days, 0 h and 15 minutes ago (should be days/weeks)