kiwamu.izumi@LIGO.ORG - posted 20:13, Monday 08 December 2014 - last comment - 10:44, Wednesday 10 December 2014(15499)
HAM2 ISI sensor correction turned off

Evan, Krishna, Alexa, Dan, Kiwamu,

At some point in this evening, we noticed that the IMC kept dropping its lock for some unknown reason. It turned out that the HAM2 ISI sensor correction was amplifying seismic at 0.3-ish Hz which resulted in a large drive in MC2 to keep it locked. So we turned the sensor correction off for now. This fixed the issue.

Even though the MC2 coil DACs were not saturating, somehow the motion was big enough to unlock the IMC. It is still unclear why it could unlock. Anyway, after turning off the sensor correction, the amont of drive in MC2 suspension reduced by a factor of 5 or so and IMC became able to stay locked. With the sensor correction on, the MC length was moving approximately +/- 5 um according to IMC_X displacement monitors. We don't know what changed in the HAM2 ISI sensor correction as it has been running fine for a while recently. Only thing we immidiately noticed was that the seismic freq-limited RMSs were pretty high in 0.1-0.3 Hz and 0.03-0.1 Hz bands in this evening. The attached is a time series of the IMC_X signals when we did a simple test of turning on and off the sensor correction in the HAM2 ISI. Since we were a bit rough for turning on and off the correction, it gave a transient when switcing it, but one can still see that when the correction was running, the IMC_X signal increased by roughly a factor of 5.

Images attached to this report
Comments related to this report
evan.hall@LIGO.ORG - 19:42, Tuesday 09 December 2014 (15522)

According to Valera, we should have sensor correction on for both HAM2 and HAM3, or for neither.

If both of them have sensor correction on, the MC benefits from the reduced seismic noise. If both of them have sensor correction off, the MC benefits from some amount of common-mode rejection in the ground motion between HAM2 and HAM3. If only one or the other has sensor correction on, then the MC is fully exposed to the ground motion, since one platform moves with the ground and the other is isolated.

valery.frolov@LIGO.ORG - 10:44, Wednesday 10 December 2014 (15532)
For PRC/SRC the sensor correction has to be ON on all HAMs (as BSCs are already inertial due to low blend) to avoid the full ground motion impression on these cavities.