R. Lane, A. Ramirez Testing ITMY (QUAD 2) Took the Top Mass OSEM Coil Resistances, Photo Diode Voltages, and LED Voltages today. Took measurements at the connection between the octopus cables (D1000234) and the top mass extension cable. The OSEMS meet the requirements.* Coil Resistances: 39±1Ω PD Voltages: 1.01±0.05V LED Voltages: 0.53±0.05V Main Chain OSEM S/N Cable Coil(Ω) LED(V) PD(V) Face 1 481 695 36.3 1.017 0.554 Face 2 638 695 36.0 1.018 0.551 Face 3 501 695 36.4 1.018 0.558 Left 496 695 36.0 1.017 0.551 Right 499 698 36.5 1.018 0.558 Side 445 698 36.7 1.017 0.558 Reaction Chain OSEM S/N Cable Coil(Ω) LED(V) PD(V) Face 1 471 698 35.9 1.018 0.553 Face 2 484 698 36.4 1.018 0.554 Face 3 472 707 36.9 1.016 0.554 Left 630 707 36.8 1.017 0.562 Right 466 707 36.7 1.019 0.558 Side 415 707 36.0 1.018 0.559 *The requirement for the coil resistance to be 39±1Ω was defined at LLO using "extra long" non final cables. This requirement is not sufficient to determine if the resistance is within standard. I will inquire with the correct authorities to determine what the new value should be for the in-vacuum final cables. However, they do meet the tolerance of ±1Ω. We took the measurement at the octopus cable, not at the vacuum feed through simulator as specified. We did not have the cables available. Although this should not add more than a few tenths of an ohm to the final resistance and would not make up the difference.