Previously, we had no guidelines on how to connect our 25-pin cables (they are used for the Trilliums [D1100152] and as Extensions for the GS13s/L4Cs [D1100153]). These cables have Peek connectors on both ends, but one end is nickel plated; we weren't aware of an orientation for these cables, so we randomly had Peek/nickel plated connectors connected at the feedthru.
We were notified that the Peek connector must be connected at the feed-thru. This morning (4) of our cables were flipped around for the BSC8-ISI. BSC6-ISI is also connected correctly as well. Attached are photos of the wrong connectors plugged in at the feed-thru and the right ones plugged in.