Laser Status: 
SysStat is good
Output power is 29.3 W (should be around 30 W)
FRONTEND WATCH is GREEN
HPO WATCH is RED

PMC:
It has been locked 4day, 22hr 0minutes (should be days/weeks)
Reflected power is 2.9 Watts  and PowerSum = 25.8 Watts.
(Reflected Power should be <= 10% of PowerSum)

FSS:
It has been locked for 0h and 35 min (should be days/weeks)
Threshold on transmitted photo-detector PD = 1.74V (should be 0.9V)
ISS:
The diffracted power is around 7% (should be 5-15%)
Last saturation event was  h and 16minutes ago (should be days/weeks)

There was a group of four working in the enclosure today taking RFAM measurements causing some grief to the normal stability of things.