Reports until 09:15, Thursday 20 October 2011
H1 PSL
lutz.winkelmann@LIGO.ORG - posted 09:15, Thursday 20 October 2011 (1592)
Pump currents for PSL high power oscillator

Measured pump currents for different output power levels at the individual fiberbundle tip for the high power oscillator of the PSL.

The output power was measured with a thermopile 300 W external OEM Ophir power meter.

 

output power at finberbundle tip / W pump current FB 1 / A pump current FB 2 / A pump current FB 3 / A pump current FB 4 / A
         
10 11.1 11.4 11.4 11.4
20 12.8 13.0 13.1 13.0
30 14.5 14.6 14.7 14.7
40 16.2 16.2 16.3 16.3
50 17.9 17.8 17.9 18.0
60 19.6 19.4 19.6 19.6
70 21.3 21.0 21.2 21.2
80 23.0 22.6 22.8 22.9
90 24.7 24.2 24.4 24.5
100 26.4 25.8 26.1 26.2
110 28.1 27.4 27.7 27.8
120 29.8 29.0 29.3 29.4
130 31.5 30.6 30.9 31.1
140 33.2 32.2 32.6 32.7
150 34.9 33.7 34.2 34.4
160 36.6 35.3 35.8 36.0
170 38.2 36.9 37.4 37.6
180 39.9 38.5 39.1 39.3
190 41.6 40.1 40.7 40.9
200 43.3 41.7 42.3 42.6
210 45.0 43.3 43.9 44.2
220 46.7 44.9 45.6 45.8
230 48.4 46.5 47.2 47.5
240 50.1 48.1 48.8 49.1
250 51.8 49.7 50.4 50.8
260 53.5 51.3 52.1 52.4
270 55.2 52.9 53.7 54.0
280 56.9 54.5 55.3 55.7
290 58.6 56.1 56.9 57.3
300 60.3 57.7 58.6 59.0