Displaying report 1-1 of 1.
Reports until 06:57, Thursday 08 January 2015
H1 SEI (SEI)
fabrice.matichard@LIGO.ORG - posted 06:57, Thursday 08 January 2015 - last comment - 14:33, Thursday 08 January 2015(15933)
HAM3 0.6 Hz noise

Rich, Jim, Seb, Fabrice

 

We can only see this noise line when the sensor correction is ON, so we have been chaising it in the the sensor correction channels, but could not find anything.

We are back to noise hunting in the ISI channels.

- The first plot attached shows  ASDs with sensor correction ON (2 nights ago). The amplitude of the peak is 2.5e-8 m/Hz^0.5

- The second plot shows ASDs with sensor correction OFF (last night).  The amplitude is 8e-9 m/Hz^0.5. No visible noise line...

 

We are getting the ISI local sensors data on HAM2 and HAM3 with the sensor correction OFF,  to perform signal subtraction, and see if the noise line could be burried in the seismic signal (and somehow amplified by the sensor correction?)

Images attached to this report
Comments related to this report
fabrice.matichard@LIGO.ORG - 14:33, Thursday 08 January 2015 (15949)

I did the subtraction between HAM2 and HAM3 ISI channels.  I have attached the cartesian results for Z, which was the most interesting. The first plot attached is with sensor correction ON. The second plot attached is with sensor correction OFF.  The residual with sensor correction OFF shows nothing at 0.66 Hz.  

 

It might worth to have another look at the coherence between the output of the FIR filters of HAM2 and HAM3 (someone one site will have to do it, I can't access those remotely, thanks)

 

Images attached to this comment
Displaying report 1-1 of 1.