Reports until 09:41, Saturday 10 January 2015
H1 CDS (CAL)
david.barker@LIGO.ORG - posted 09:41, Saturday 10 January 2015 (15994)
Beckhoff restart of h1ecaty1 for PCAL binary output switching

Rick, Sudarshan, Shivaraj, Dave, Daniel:

late entry from Friday morning. Daniel found the PCAL binary output switching of the servo, laser power and shutter for PCAL ETMY was due to a mis-configuration within h1ecaty1. He fixed the configuration and restarted all three PLCs on h1ecaty1. He then burt restored these IOCs.