PCal EY Status

Sudarshan, Rick, Darkhan

Pcal EY is running at full power with 2W beam pushing the end test mass and Optical Follower servo engaged. The loop gain is set at 24dB with refrence offset at 5 V. We will let it run over the weekend and characterize the PDs on Monday.