Reports until 08:16, Wednesday 28 January 2015
H1 AOS
douglas.cook@LIGO.ORG - posted 08:16, Wednesday 28 January 2015 (16321)
ITMy OPLVR being scrubbed by copper pipes
I opened the gap between the bundle of copper pipes that were again rubbing on the ITMy optical lever TX pier causing the 10min-ish sawtooth signal. This is a temporary fix, but is more stable than the first time we saw this. It is in the planning to reroute the pipes and fix other similar issues with other oplvrs.
Keita ran a new scan verifying the fix.