The LSC model has been edited to accommodate a normalized REFLAIR9I signal (LSC-TR_REFLAIR9) alongisde normalized REFL9I (LSC-TR_REFL9). A switch determines which of these is signals is fed into LSC-REFLBIAS.
Correspondingly, I have switched back the cabling on the ISC rack so that the demodulated REFL_A_RF9 signals feed into the REFL_A_RF9 whitening chassis, and likewise for REFLAIR_A_RF9.
The guardian scripts, whitening settings, and demod phases are now corrected back to their settings prior to the cable swap.