H1 SEI (DetChar)
jeffrey.kissel@LIGO.ORG - posted 15:53, Monday 02 February 2015 - last comment - 17:22, Monday 02 February 2015(16419)
Field HPI Pump Servo Sensor Inputs Terminated
J. Kissel, H. Radkins

Hugh and I continue to investigate the corner station HEPI pump servo noise at its source. As such, we've switched the servo OFF and terminated the two pressure sensors used for the loop error point (read in as H1:HPI-PUMP_L0_BSC2SUP_PRESS, and H1:HPI-PUMP_L0_BSC2RET_PRESS) with 100 [ohms] at 3:45 pm (23:45 UTC). The hope is to quantify the servo's ADC noise floor. We intend to leave it in this configuration for an hour or so before restoring it. Meanwhile, Hugh continues to manually maintain the pump pressure (though very little has been required but for a few 1 [ct] tweaks).
Comments related to this report
jeffrey.kissel@LIGO.ORG - 17:22, Monday 02 February 2015 (16421)DetChar
Pressure sensors have been reconnected at 2015-02-02 5:01pm PST (2015-02-03 01:01 UTC), and pump servo turned back ON at 5:20pm PST (01:20 UTC).