Reports until 10:10, Thursday 05 February 2015
H1 AOS
eleanor.king@LIGO.ORG - posted 10:10, Thursday 05 February 2015 - last comment - 13:50, Thursday 05 February 2015(16486)
End-Y PLL fiber polarization changes

Kiwamu, TJ, Elli

This morning while EY plugging HWS cables into the ISCTEY feedthrough panel, I touched the fiber bringing the red PLL reference  signal to EY.    This changed the fiber wrong polarization (see chanel H1:ALS-Y_FIBR_LOCK_FIBER_POLARIZATIONPERCENT) from 1% to 32%, which is above the maximum allowable value of 30%.  This fiber is also connected to ISCTEY feedthrough panel, and then some of the fiber is sitting on a coil on top of ISCTEY.  All I did to change the polarization was to gently move the coil on top of the table.  When Kiwamu and TJ noticed the polarization had changed, I nudged the fiber coil on top of the table around a bit untill the fiber polarization returned to 8%. It seems it is very easy to bump the fiber and change the polarization.

Images attached to this report
Comments related to this report
daniel.sigg@LIGO.ORG - 13:50, Thursday 05 February 2015 (16493)

One can use the fiber polarization controller in the MSR to bring it back.