Laser Status: SysStat is good (except VB program online = known issue) Output power is 32.4 W (should be around 30 W) FRONTEND WATCH is Active HPO WATCH is red PMC: It has been locked 20 day, 2 hr 23 minutes (should be days/weeks) Reflected power is 2.1 Watts and PowerSum = 25.3 Watts. (Reflected Power should be <= 10% of PowerSum) FSS: It has been locked for 7 h and 17 min (should be days/weeks) Threshold on transmitted photo-detector PD = 1.12 V (should be 0.9V) ISS: The diffracted power is around 5.4 % (should be 5-15%) Last saturation event was 7 h and 19 minutes ago (should be days/weeks)