H1 PSL (DetChar, PSL)
jason.oberling@LIGO.ORG - posted 14:36, Tuesday 10 February 2015 - last comment - 09:48, Wednesday 11 February 2015(16605)
PSL Maintenance Summary

Rick S., Jason O., Matt H., Ed M., and Jeff B.

Went into the PSL today and performed a few maintenance tasks (work permit #5039).

1) Measured PSL power at several points in the beam path:

2) Measured power in FSS path before RefCav:
Note: Make sure ISS is off. Make note of ISS offset (4.2%) and diffracted power (5.2%). Check when finished.

3) Measured voltage of RefCav REFL PD (H1:PSL-FSS_RFPD_DC_OUTPUT):

Once again the RefCav Trans PD (TPD) has begun to drift down.  Was set to ~1.6V on 1/5/2015 (see alog 15871) and today is reading ~0.9V.  Therefore we adjusted the RefCav alignment by adjusting the vertical and horizontal of the top periscope mirror.  Measured voltages at the RefCav REFL PD.

After adjustment (and locking of adjustment screws):

Will keep an eye on this as the RefCav transmitted power seems to drift down suddenly.  We’re not sure what’s causing this apparent alignment drift.  All the measured powers leading up to the RefCav were close to those measured during the 1/5/2015 adjustment, but the RefCav TPD was still reading ~44% less.

4) We also measured the UGF of the FSS:

5) For Daniel, we adjusted power at IO PD (IO-AB-PD3-DC) to 3.3V by turning up the PD gain by 3 clicks and adding a ND filter to the PD (see Daniel's earlier alog).

6) We are now storing the 2 3IFO PMCs in the NW corner of the PSL LAE on a cart, with their lids off.

Comments related to this report
matthew.heintze@LIGO.ORG - 16:51, Tuesday 10 February 2015 (16609)

Now that have a measured power budget here at LHO for some areas of the PSL table (note for completeness that the power readings at LHO were done using 250W water cooled meter and at LLO using 50W air cooled meter) we can compare the losses through the system to (and after) the PMC. 

 

Position LHO LLO
(1) Out of MOPA (35W laser) 33W 34.25W (alog 16569)
(2) After faraday in HPO N/A 31.44W (alog 16569)
Delta (1) - (2) N/A 2.81W
(3) At HPO window with corona aperture out N/A 29.1W (alog 16569)
Delta (1) - (3) N/A 5.15W
(4) In front of PMC (ISS off) 27.2W 28.75W (alog 16569)
Delta (1) - (4) 5.8W 5.5W
(5) Power Trans (power out of PMC) 24.9 W (ISS off) 26.9 W (from todays monitor screen (ISS looks to be on))
Throughput ((5)/(4)) 91.5% ~93.5% (probably slightly better if ISS was off as (5) would be slightly higher)
Visibility   93% (alog 16576)
Delta (1) - (5) 8.1W 7.35W
(6) After EOM (ISS off) 24.4W 25.7W (alog 16576)

 

So basically the performance of the two laser systems are similar. LLO has slightly less loss from the output of the MOPA to the input of the PMC (5.5W loss at LLO compared to 5.8W at LHO). Also LLO has slighlty less loss of power once comes out of the PMC as well (LLO ~7.35W dropped from output of MOPA to output of PMC, compared to 8.1W at LHO). But all in all the two systems in terms of loss through the PSL components is very similar

jason.oberling@LIGO.ORG - 09:48, Wednesday 11 February 2015 (16644)

Here is a 60 day trend of the FSS RefCav transmission (H1:PSL-FSS_TPD_DC_OUT_DQ) showing the two drops in PD voltage we've seen in the last couple months.  The first occurred around 12/25/2014 and was adjusted on 1/5/2015.  The latest drop happened around the middle of last week and we adjusted it yesterday.