CDS Maintenance Summary

TCS slow channels

Alastair, Aidan, Dave. WP5043

The two laser setpoint channels which were temporarily added to script0's IOC last friday were removed and replaced by these plus more chanels in the h1tcscs front end model. The full list of new channels added to tcs/common/model/TCS_MASTER.mdl are

TCS_MASTER.mdl submitted to SVN r9793. This closes WP5043

PEM DAQ channels

Peter F, Robert S, Dave. WP5042

The PEM models h1pemcs, h1pemex, h1pemey, h1pemmx and h1pemmy were modified to conform with Peter and Roberts PEM data rates to frame document T1400768. There remains an inconsistency in the number of magnetometers and accelerometers, so this WP will remain open. Code was committed to SVN r9794

PCAL addition of working standard channel

Rick, Sudarshan, Dave. WP5045

An additional ADC channel was added to the PCAL model to read out the Working Standard signal. The model files pem/h1/models/h1pem[ex,ey] and cal/common/models/PCAL_MASTER.mdl were modifed. The new code went in at the same time as the above PEM restarts because CAL and PEM were merged. The PCAL_MASTER.mdl was committed to SVN r9763. This closes WP5045.

DMT Broadcaster and DAQ configuration

John Z, Dave

The following channels were added to H1BROADCAST.0.ini

The ALS testpoint par file was added to the DAQ (was accidentally left out during the ISC split last week)

DAQ Restart

Jim B, Dave

To support all the above, the DAQ was restarted. Jim delayed the startup of h1fw1 and h11nds1 to reconfigure their NFS mounts to support minute trend offloading on h1fw1.

Cleanup

Dave

I pressed all the DIAG_RESET buttons to clear accumulated ADC,IPC,etc. errors.

The following models had modified filter file warnings, I did a full filter file load on them: h1susmc2, h1iscex, h1iscey.

h1susmc2 did not seem to actually have a modified file (filter_archive files were identical).

isc end stations were a residual of the ISC split last week, so of the filters left in the isc models no actual filter changes were applied.