Reports until 00:14, Thursday 12 February 2015
H1 ISC
daniel.hoak@LIGO.ORG - posted 00:14, Thursday 12 February 2015 (16678)
DCPD balancing

Today I measured the balancing of the OMC DCPDs, using an excitation at 103Hz into the OMC length servo, and measuring the height of the second harmonic in the nullstream channel.  The DCPDs are only, as far as I can tell, 1% misbalanced.  In the attached plot data are in red, the blue is a by-eye fit.  I have set the balance to be 50.5% PDA, 49.5% PDB (1% towards PDA on the slider).

Non-image files attached to this report