I reset the dust monitor counts alarm levels to reflect the US Fed 209E cleanroom standards, as defined in T1400024.
For Clean-1000 (General room monitoring of LVEA and VEAs)
0.3 minor 2000, major 3000
0.5 minor 750, major 1000
For Clean-100 (Open chambers, cleanrooms with Class-A, etc.)
0.3 minor 250, major 300
0.5 minor 75, major 100
The dust monitor alarm have been set to these levels. I will monitor over the next couple of weeks to see if these levels need to be adjusted.