Reports until 18:12, Monday 02 March 2015
H1 ISC
sheila.dwyer@LIGO.ORG - posted 18:12, Monday 02 March 2015 (17028)
DRMI + arms off resonance ASC

Evan, Alexa, Gabriele, Sheila

We have closed 8 DRMI ASC loops with the arms off resonance.  Ordered fast to slow they are:

These loops are now all turned on by the guardian, mich comes on first, there is a pause and the rest come on.  We can leave these on for the first steps of the CARM offset reduction.  We have manually been turning off all of them except for MICH at this point.  We tried leaving the PRM loop on once, this caused bad drift without the 2 refl loops closed. 

WARNING:  the guardian turns on 6 loops durring the DRMI ASC step that need to be turned off manually before the CARM offset is reduced too much

Images attached to this report