Reports until 00:03, Wednesday 04 March 2015
H1 AOS
sheila.dwyer@LIGO.ORG - posted 00:03, Wednesday 04 March 2015 - last comment - 09:18, Wednesday 04 March 2015(17061)
Comm beatnote

After maintence day this morning, the corner beckhoff was not restored.  The ALS SHG was not working because the temperature was not controlled.  Before we realized this we went to the table to investigate and tried to adjust the beat note alingment.  Once the SHG was back on, the beat note stregth was low.  Although this sounds impossible, the thing that seemed to fix the problem was unplugging the RF amplifier and plugging it back in.  

Comments related to this report
daniel.sigg@LIGO.ORG - 09:18, Wednesday 04 March 2015 (17068)

Looking at the code I see that the save/restore was disabled for the TEC controller. Not sure why. Fix it (in svn) assuimg it was mistake.