Chris, Alexa
The first two attachments show the OLTFs of the various length DOFs. In these plots, the model is the blue trace and the red trace is measured data. The DARM TF agrees pretty well. The PRCL and SRCL loops appear to have a gain missing. The CARM and MICH model loops are not correct.
The fourth attachment depicts the noise budget produced by the model for the 8W lock on March 04, 2015 15:30 UTC. This roughly agrees with the conclusions Evan has made with his NB (LHO#17101). At high frequencies we are dominated by shot noise. At around 300 Hz we suspect intensity noise (or maybe beam jitter), as Gabriele noticed. From 30-100 Hz we are limited by DAC noise. And below 10 Hz we see some angular fluctuations.
The modeled MICH loop gain is looking better now. There were some aggressive BSFM plant inversion filters that weren't being linearized accurately in Simulink. This was fixed by using numerical TFs for them. The shape is still off above 60 Hz, but that's probably due to the filter change from alog 17148.
I put in sign flips as needed to ensure an even number of them are present in each OLTF. There remains some 10 dB of mystery gain to be hunted for in the DRMI loops.