Due to all of the latest commissioning of ETMY, the SDF had accumulated 40+ differences. With Jeff's help we remedied numerous line items via making the ETMy ESD banks look like the ETMx ones. This cleared out Kiwamu's settings leftover from earlier measurements. I've also moved the ESD calibration filters from the 4 ETMY L3 ESD OUTPUT banks to the L3 LOCKING BIAS bank to bring it in line with the ETMx settings. I've edited the burt ETMy safe.snap to capture some of the other commissioning work. THere are now only 20 SDF diffs which I will continue to track down.
Screen snapshots before ETMy cleanup:
ETMy screen snapshots now:
Betsy, Jeff,
Thanks for cleaning up the mess and I am sorry that I did not alog the details. Regarding this configuration, I actually liked the ETMY configuration than ETMX because (1) the effective charge can be calibrated in Volts at the electrodes and (2) the bias voltage can be specified on the medm screen as the actual voltage at the electrode e.g. +380 V. We should decide which calibration convention we want to use. I will talk to you guys about this once I come back.