Dan, Kiwamu, Sheila
Today we made several gaurdian changes:
I looked for a signal from the QPDs which was insensitive to TMS angle. For the X arm pitch I used TRX A - 0.5 TRB and for Y arm pitch 1TRY_A-0.65 TRY_B seemed good. I closed a verry low bandwidth loop to ITMX using this signal, screen shot is attached.
We found that we could increase the recycling gain from ~27 to ~33 by moving the ITMs with ASC lops closed on IM4, PR2, BS, common ETMs, and differential ETMs. We had to move ITMX by as much as 5 urad to do this, and we found that we would loose the lock soon after getting a good recycling gain. Kiwamu calcualted that by moving ITMX 5 urad, we are moving the spot on the test masses by 1 cm assuming that the ETM is controlled. This may be due to a problem with the DHARD loop.
We also spent some time looking for a Transmon QPD signal that would be insensitive to motion of the transmon. We closed a loop to ITMX pitch using TRX_A - 0.5 TRX_B, the attached screenshot shows the settings. This loop was verry low bandwidth.