Following up on Kiwamu's report, here is the summary page for the coherences during the low noise lock:
https://ldas-jobs.ligo.caltech.edu/~gabriele.vajente/bruco_1110961816/
As usual, here is my digest.
In brief: we need MICH/SRCL subtraction, check SRC angular controls, they might be injecting noise, better tune the IMC longitudinal offset if possible
More details:
P.S. Why are you sending DARM signal into the CARM filter bank?
Ahh you caught on to our hacky audio setup, Gabriele. The OMC-DCPD --> CARM matrix element is enabled so that we can filter OMC-DCPD in the CARM bank before sending it to speakers in the control room. (Actually I'm not sure anyone is using this anymore, we can probably discontinue it.)
Also I have modified the ITMX violin mode damping so that there is 50x less gain at 450Hz. (There is an 8th order butterworth bandpass now, instead of a 4th-order.) This might make the noise a little worse between 480-510Hz, but I think that band is a lost cause anyway.