Reports until 18:18, Thursday 26 March 2015
H1 SUS (CDS, ISC, SUS)
evan.hall@LIGO.ORG - posted 18:18, Thursday 26 March 2015 - last comment - 07:00, Friday 27 March 2015(17496)
ETMY ESD driver tripping repeatedly

Rich A., Lisa, Sheila, Jeff, Evan

The EY ESD driver has tripped twice today, apparently in the same failure mode as described previously. Each time we have had to drive down to the end station to reactivate it.

The supplies are set at ±430 V, each with a 75 mA current limit. When the driver is active, it draws 23 mA dc from the positive supply and 20 mA dc from the negative supply. When tripped, it draws 3 mA dc from each supply.

Comments related to this report
sheila.dwyer@LIGO.ORG - 00:55, Friday 27 March 2015 (17500)

This seems to happen every time we loose lock if we have transitioned to ETMY.  We are getting plenty of excerise tonight going back and forth to the end station.  

If anyone has any ideas about how to make this easier to reset or to prevent it from happening, that would save us some fuel.  

kiwamu.izumi@LIGO.ORG - 07:00, Friday 27 March 2015 (17508)

Why don't we set a digital limiter ? For example, see LLO alog:

https://alog.ligo-la.caltech.edu/aLOG/index.php?callRep=14472