Reports until 09:35, Monday 30 March 2015
H1 PSL
edmond.merilh@LIGO.ORG - posted 09:35, Monday 30 March 2015 (17566)
PSL ISS and FSS

I took a look at the ISS this morning and found the dfifracted power up around ~11% + with a ref signal of ~ -1.8. It's been adjusted to ~ 7% at a refsignal of -2.02. Also, FSS TPD[V] are down to 1.4 from 1.6. There maybe the need for incursion on maintenance day to do touch ups and discovery.