102.4kHz bandwidth amplitude spectral density of OMC DCPD 1 output, Ed Daw, Ross Kennedy and Daniel Hoak
Here are results from an SR785 spectrum of the output of the DCPD whitening field amp for OMC DCPD 1. I have included the ASD plot as a PDF and listed the major frequencies below. Below Nyquist the frequencies are consistent with H1:IOP-LSC0_MADC0_TP_CH12, so I have used these instead of the frequencies derived from the SR785 data. The amplitudes are from the SR785 data throughout. Those amplitudes differ from H1:IOP-LSC0_MADC0_TP_CH12 as they are measured before the whitening filter / antiwhitening DSP.

Index. Frequency (Hz) ; Amplitude (uVrms/rtHz)
1. 8160 Hz ; 595 uV
2. 9880 Hz ; 177 uV
3. 10426 Hz ; 206 uV
4. 13000 Hz ; 135 uV
5. 13857 Hz ; 145 uV  (wideband cluster in H1:IOP-LSC0_MADC0_TP_CH12 - looks narrower in SR785 data)
6. 15072 Hz, 15218Hz ; 136uV (double narrow peak in H1:IOP-LSC0_MADC0_TP_CH1 - unresolved in SR785 data)
7. 19554 Hz ; 52 uV
8. 21128 Hz ; 23 uV
9. 23109 Hz ; 22 uV
10. 25000 Hz ; 21 uV
11. 26607 Hz ; 19 uV
12. 29769 Hz ; 25 uV
13. 32693 Hz ; 47 uV (wideband, but not resolved well on SR785)
14. 35775-38594Hz ; peak 87 uV (wideband collection of unresolved lines in SR785)
15. 39964 Hz ; 27 uV
16. 42467 Hz ; 22 uV
17. 47287 Hz ; 5.2uV
18. 47762 Hz; 3.6uV
19. 75107 Hz ; 13uV (very large peak relative to surrounding background)
20. 98000 Hz ; 3.2uV
21. 101630 Hz ; 21uV (very large peak relative to surrounding background)
Non-image files attached to this report